6–4
If 〈NO/CE〉 is pressed from the Want to view...? screen on the previous
page, the display is:
Put gauge on Std
Block, rod in
SAFE position &
press START
For a stat test, place the reference standard block on a high-density surface
(at least 100 pcf (1602 kg/m
3
)) that is 5 ft (1.5 m) from a vertical surface.
Following the instructions given on the display, place the gauge on the
reference standard block with the source rod in the SAFE position.
Press 〈START/ENTER〉 for the display:
- STAT TEST -
Reading #1
Time= 60 secs
The gauge will take twenty 60-second counts. When all counts are
complete, the screen will display the average counts and ratio.
DENS STAT TEST
Avg cnts: xx
R: x.xxx
-ENTER for Moist
To view the moisture data, press 〈START/ENTER〉. The display is:
MOIST STAT TEST
Avg cnts: xx
R: xxxx
View Stat. data?
Press 〈YES〉 for the display:
Select method of
viewing data:
1- Screen
2- Printout