Model 3440-L 7–5
7. EXTENDED FUNCTIONS
CALIBRATION DATE
The Calibration Date function allows the operator to change the date of the
last gauge re-calibration. To execute the Calibration Date function, press
SHIFT and SPECIAL for:
SPECIAL FUNCTION
YES- Next menu
1- STAT TEST
2- DRIFT TEST
Press 19 for the display:
EXTENDED
FUNCTIONS
Enter Code ----
and Press ENTER
The Access Code must be entered. This code is for authorized personnel
only. If you need assistance with the code, contact your nearest Troxler
Representative.
Enter the code and press START/ENTER. The display will be:
YES- Next menu
1- Time/Date
2- Customer Name
3- Serial Number
Press YES for the following display:
YES- Next menu
4- Calib. Date
5- Calib. Const.
6- Rod Length
Press 4 for the display:
DATE OF Calib:
--------
Press ENTER
when complete.
Using the numeric keys, input the new date. Press START/ENTER
when completed. The display will return to the Extended Functions menu.