EasyManua.ls Logo

Troxler 3440-L - Stat Test

Troxler 3440-L
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 3440-L 6–3
6. SPECIAL FUNCTIONS
STAT TEST
The Stat Test, or statistical stability test, may be performed to validate the
normal operation of the gauge. Erratic readings or readings that seem to
fluctuate may indicate a problem with the gauge. If the readings are
suspect, the stat test may be executed.
A stat test should result in a ratio of actual-to-predicted standard deviation.
Ideally, the ratio should be 0.25, but acceptable limits are between 0.17 and
0.33. A definite instability exists if the ratio is outside these limits. Contact
the nearest Troxler Service Center for additional instructions.
To execute the stat test, press SHIFT and SPECIAL for the display:
SPECIAL FUNCTION
YES- Next menu
1- STAT TEST
2- DRIFT TEST
Press 1 for the display:
- STAT TEST -
Want to view
last Stat data?
If YES is pressed to view the last stat test data, the display is:
DENS STAT TEST
Avg cnts: xx
R: x.xxx
-ENTER for Moist
To view the moisture data from the last stat test, press START/ENTER.
The display is:
MOIST STAT TEST
Avg cnts: xx
R: x.xxx
View Stat. data?
The data for each of the readings may be viewed on the screen as described
on page 6–5.

Table of Contents

Related product manuals