5–12
AUTO-STORE
The Auto-Store function automatically stores sample data upon
completion of a burn cycle. The data is stored under the active
project, using a sequential auto-sample ID number. When a new
project is created, the sample ID number for the Auto-Store function
starts at 1 and is incremented each time a sample is stored
automatically.
NOTE
The Auto-Store sample ID number is not incremented if
sample data is stored manually (see page 5–24). When
using the manual Store function, the operator ente
rs a
sample ID number manually. Therefore, duplicate
sample ID numbers could be stored under the acti
ve
project.
For example, if the Auto-Store function is turned on for
a new project, the oven stores the sample data for the
first two burn cycles under sample IDs 1 and 2. Assume
that the Auto-Store is then turned off, and the operator
manually stores data from two more burn cycles using
sample IDs 3 and 4. If the Auto-Store function is then
turned on, the sample data for successive burn cycles
will also be stored using sample IDs 3, 4, and so on.
To prevent duplication of sample ID numbers, Troxler
recommends either using sample IDs beginning at 100
or higher when storing sample data manually or always
enabling the Auto-Store function.
To access the Auto-Store function, press 〈3〉 at the Burn Setup
Menu. The oven displays:
Auto-Store
1. On
2. Off
f To return to the Burn Setup Menu without changing the
Auto-Store status, press 〈ESC〉.