6–8
CORRECTION MENU
The oven can store up to twenty aggregate correction factors
(ACFs). Each ACF is stored using an alphanumeric ID entered by
the operator. The ACF ID can be up to twelve characters. When a
correction factor is active, all measurements are adjusted using that
correction factor.
The Correction Menu enables the operator to select (activate) an
existing ACF, to create a new ACF, or to erase one or more stored
ACFs. To access the Correction Menu, press 〈4〉 at the MAIN
MENU. The oven displays:
Correction Menu
1. Select ACF
2. Create ACF
3. Erase ACF
To select an item from the menu, press the corresponding number
key.
SELECT ACF
To select (activate) an existing ACF, press 〈1〉 at the Correction
Menu shown on page 6–8. If there are no stored ACFs, the oven
displays the error mess
age No ACFs Found and returns to the
Correction Menu.
If there are stored ACFs, the oven displays:
Select
1. Scroll ACFs
2. Enter ACF ID