APX 525/625 Generator Revision C Installation
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 100
INSTALLATION
RACK MOUNTED APX 525/625 X-RAY GENERATOR
Warning: The main SCR capacitor bank (bottom of power module) contains a very
high charge when power is applied. This charge is a fatal shock hazard. After power
has been disconnected from the system, either by line disconnect or control console
power off, allow a minimum of five minutes for capacitor bank bleed down. Check
capacitor bank for zero volts with a dc voltmeter before working on any internal
circuitry.
Caution: Generator performance will be impaired if additional circuitry is powered by
the same incoming line. There are no accessory power provisions within the
generator. Attempts to attach accessory items such as collimator or table power
sources to this generator will damage circuitry.
Note: The UNIVERSAL APX 525/625 X-ray generator requires a dedicated input line.