APX 525/625 Generator Revision C Using Del APR Utility
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 194
Error Messages
Error Message Cause & Correction
Data Mismatch - Aborted Happens when you send database with
the wrong X-Ray tube selected. Reselect
X-Ray tube and resend.
Errors in Database - Aborted Happens when you send a database with
unallowable settings. Unallowable
settings will be indicated on screen when
red dots appear next to settings or menu
cells are shaded red. Click Cancel to
continue.
Error! CRC or timeout. Happens when you send data during the
console flash update. This will have no
affect on sending the database. Continue
as normal.