APX 525/625 Generator Revision C Introduction
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 3
the generator. Tube inserts of load capacities other than those for which the APX
525 or 625 was configured may be permanently damaged if used with the incorrect
tube loading program. Consult Del Medical Imaging Corporation for additional
information. A list of compatible tubes is on page 128.
Features and Benefits of the APX 525 and 625
The advanced APX family of generators utilizes the latest in multi-pulse high
frequency technology to produce waveforms near constant potential. The features
include:
• Menu driven, soft button approach makes the APX very easy to learn
and gives maximum user flexibility
• Microprocessor control improves ease of use, reliability and
serviceability
• Two point technique selection mode utilizes Isowatt tube current
control, which provides the maximum mA at the selected combination
of kV and mAs for the shortest possible exposure times
• Operator has the ability to exercise traditional three factor control of
kV, mA, and time, in three point operational mode
• 30 kW output with APX 525 provides up to 500 mA and 40 kW output
with APX 625 provides up to 600 mA for motion stopping power
• Back lighted liquid crystal display is easy to read from almost any
angle or lighting condition
• Optional AEC mode
• Post reading mAs (in AEC mode) lets the technologist confirm
exposure and verify technique
• Built-in anode heat calculator automatically displays percentage of
heat units on anode
Power Levels
kW Rating
30 40
kV Range
40-125 40-125
mAs Range
1-500 1-600
Maximum mA
500 600