APX 525/625 Generator Revision C Service Adjustments
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 128
Note: If your desired tube is not on the tube list, contact Del Medical Imaging
Technical Support for assistance.
X-Ray Tube Assemblies
Varian RAD 8 - 1.0 X 2.0
RAD 13 - 1.0 X 2.0
RAD 14 - 0.6 X 1.2
RAD 16 - 1.0 X 2.0
RAD 21 - 0.6 X 1.2
RAD 25 - 1.0 X 2.0
RAD 40 - 0.6 X 1.2
RAD 56 - 0.6 X 1.2
RAD 60 - 0.6 X 1.2
RAD 68 - 0.6 X 1.2
RAD 68 - 1.0 X 2.0
RAD 74 - 0.6 X 1.5
Toshiba
DR-3724H - 0.6 X 1.2
UX-51H39 - 1.0 X 2.0
UX-51H42 - 0.6 X 1.5
Eimac
A119 - 1.0 X 2.0
A132 - 0.6 X 1.2
A272 - 0.3 X 0.7
A292 - 0.6 X 1.2
Dunlee
DU303M - 1.0 X 2.0
PX-1312 - 0.6 X 1.2
PX-1429 - 0.6 X 1.2
PX-1436 - 0.6 X 1.2
Comet
DX39H - 0.6 X 1.5
5. Press <exit> key to save changes and allow control to finish the power up
sequence
6. Turn generator power OFF before beginning the next step