APX 525/625 Generator Revision C Service Adjustments
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 127
The installer must program default system configuration in the following manner:
1. Depress and hold the second button from the lower left corner of the display
screen while turning the generator ON.
2. Enter the Dealer Password, using the buttons above the display to highlight the
characters, and by adjusting their numeric values with the rotational knob.
3. The default startup screen will appear as the generator is in its power up
sequence.
4. You may now select the default settings for the following parameters:
Tomo Install (Yes or No)
Tomo Type
Tube Select (Use Rotational Knob to Select. List of Tubes is on Next
Page)
Misc Options
Flash Options
Change Password
AEC Installed
APR Updates (enabled or not enabled)
Power Level
Power Derating
At this step, you must decide if the APX 525/625 Generator will operate at full
rated output. If the unit is to be installed on a less than ideal power line, it will
be necessary to derate the output power. Pressing the Power % Level button
will toggle through the available range of derating percentages.
Derating the generator will reduce the mA output during the exposure in the 2
point mode. The effect will be longer exposure times. kV will not be affected.
In the 3 point mode, derating will lower the allowable kilowatt product, which
will limit the operator’s available technique selections. When a satisfactory
power line is connected to the generator, the programmed power level can be
increased accordingly.