APX 525/625 Generator Revision C Pre-Installation
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 29
Grounding
A separate #6 AWG wire to earth ground must be provided. Follow guidelines
established by the National Electrical Code.
Single Phase Installations (APX-525 Only)
Single phase versions of the APX-525 may be operated on the following incoming
line voltages only:
208 or 240 VAC, single phase, with one safety ground.
The part number of the single phase line matching transformer that must be installed
with these generators is: 636-0021P2. This transformer is included in kit # 112-
0612G4.
Schematic – 636-0021P2 – Single Phase Line Matching Autoformer
636-0021P2 – Single Phase Line Matching Autoformer