APX 525/625 Generator Revision C Using Del APR Utility
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 193
Uploading Modified APR Files
After modification, the optimized APR database can be uploaded to the Operator’s
Control Console in the same manner as the download process, by utilizing the “Send
Data” button on the APR utility screen.
NOTE:
The database downloaded from any generator contains the specific
signature of that generator’s power limit, tube type, and derating status.
As a result, the database may only be uploaded to a generator equipped
with the same tube, generator power limit, and derating status.
• The Operator’s Console must be in the Laptop Transfer mode, with the
“WAITING FOR CMD” message displayed.
• Click on the “Send Data” button in the Data Transfer portion of the APR utility
screen.
• The APR utility program will establish communication with the console,
transfer data, verify the checksum, and wait for the flash memory to be
updated. When the process is complete, the APR utility will display the “Flash
Updated” message and the operator’s console will return to the waiting for
command state.
• You may now exit the utility screens on the operator’s console by pressing
<Exit> at all prompts. The operator’s console will then complete the boot
sequence and power up the generator.
The upload of APR data is now complete.