Three Phase Installations
All 30 kW, three phase versions of the APX-525 generators may be operated on
the following incoming line voltages:
208, 230, 240, 380, or 480 VAC, three phase, with one safety ground and no
neutral.
All 40 kW, three phase versions of the APX625 generators may be operated on
the following incoming line voltages:
240, 260, or 280 VAC, three phase, with one safety ground and no neutral.
The part number of the three phase line matching transformer that must be installed
with these generators is: 636-5010P1. This transformer is included in kit # 112-
0612G15. Information about this kit starts on the next page.
40 kW, three phase versions of the APX625 generators may also be operated on
208 VAC, three phase, with one safety ground and no neutral. However, this
power configuration requires a special line matching transformer. The part number
for this transformer is 636-0070P1. This transformer is included in kit #112-
0612G11. Information about this kit starts on page 53.