APX 525/625 Generator Revision C Installation
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 102
Plug Connections to Power Module
J13 – Connects to operator console CPU board at J5.
Stator Connections
TB3 - 1 Phase (08)
TB3 - 2 Main (07)
TB3 - 3 Common (09)
TB3 - 4 Thermal Switch *
TB3 - 5 Thermal Switch Common *
*An isolated, normally closed set of contacts, mounted on the tube housing, which
opens if the housing is beyond normal operating temperatures.
If no housing thermal switch is present, a jumper must be placed across TB3-4 and
TB3-5 to allow an exposure.
Wall Image Receptor Connections
Connection Signal
Progeny
L/F 8000
L/F 9000
L/F 8000
(older models)
Villa
TB4-1
Wall Select 117VAC
(Switched)
3 B3 7
TB4-2 Wall Feedback 2 B2 12
TB4-3 117 VAC L B8 3
TB4-4 Signal GND 1 B1 11
TB4-5 117 VAC Return N B4 1
Chassis
GND
Chassis GND GND GND
Chassis GND
Note: For 120 VAC operation the Villa Bucky requires a jumper to be installed
between terminals 1 and 6).
If installing a grid cabinet or wall cassette holder, or if no wall image receptor will be
employed, you must place a jumper between TB4-4 and TB4-2 to enable exposures
when the wall device is selected.