APX 525/625 Generator Revision C Service Adjustments
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 129
Perform User Default Screen setup
1. Depress and hold first key from left on lower side of display window and turn
generator power ON. The console will now display the default user setup
screen.
2. Configure unit in a manner best suited for the installation :
Select a default mAs setting
Establish the 2 Point kV Default
Establish the Default Startup mode (2 or 3 point)
Establish the 3 Point kV Default setting
Select the 3 Point mA Default setting
Select the 3 Point Time Default
Select the default Image receptor (Wall, Table, or None)
Select the Display Contrast Default (Set to approximately 50)
3. Press <exit> and allow generator to complete power up sequence.
Generator is now ready for Calibration and Acceptance Testing