EasyManua.ls Logo

Universal APX 525 - Page 148

Default Icon
243 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APX 525/625 Generator Revision C Service Adjustments
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 139
Mode
3 point
Tube
1
Focal Spot
Large
kV
80
mA
200 mA
Time
100 mSec
4. Connect channel 1 to the anode BNC output of the Dynalyzer™
5. Connect channel 2 to the cathode BNC output of the Dynalyzer™
6. Set up the oscilloscope as follows:
Channel 1 Vertical Scale
2V / Division
Channel 2 Vertical Scale
2V / Division
Horizontal Scale
10 mSec / Division
Trigger
Channel 1
Polarity
Positive
Mode
Normal
7. Adjust the baseline position of the two channels so that they are both aligned on
the same centerline of the vertical scale. Channel 1 will display the positive
anode kV at 20 kV / division, and channel 2 will display the negative cathode kV
at 20 kV / division
8. Make exposure and adjust R77 on kV Board for 80 kV (+1 kV) on the
Dynalyzer™ digital display. Verify that the anode and cathode waveforms are of
nearly equal amplitudes, and that the sum of the two voltages equals 80 kV (+8
VDC differential voltage on the oscilloscope)
Figure 68 – kV Waveform

Table of Contents