APX 525/625 Generator Revision C Operating Manual
©
2005 Del Medical Imaging Corporation
7/5/2005 Page 9
Single Function Buttons
Button Description
kV
Pressing the button above this portion of the display will highlight
the current value. The kV may be changed in single digit
increments, using the rotational knob.
mAs
Pressing the button above this portion of the display will highlight
the current value. The mAs may be changed in pre-determined
steps, using the rotational knob. (2 point mode only)
mA
Pressing the button above this portion of the display will highlight
the current value. The mA may be changed between 25 mA to 500
mA (25 mA to 600 mA for 625), in 12 steps, using the rotational
knob. (3 point mode only)
Time
Pressing the button above this portion of the display will highlight
the current value. The exposure time may be changed in single
digit increments using the rotational knob. (3 point mode only)
Although values of time may be selected, the actual duration of the
exposure will always be controlled by the mAs timer circuitry. The
available Time selections (in seconds) are: .010, .012, .015, .020,
.025, .030, .035, .040, .050, .060, .070, .080, .09, 0.1, 0.12, 0.15,
0.25, 0.3, 0.35, 0.4, 0.5, 0.6, 0.75, 0.9, 1.0, 1.25, 1.5, 1.75, 2.0, 2.5,
3.0, 4.0, 5.0, and 6.0
Focal Spot
Pressing the button adjacent to this portion of the display will toggle
the current value between the large and small focal spot.
International symbols are used to depict the large and small focal
spots.
APR
Pressing the button adjacent to this portion of the display will
activate the APR mode of operation.
Image
Receptor
Pressing the button adjacent to this portion of the display will toggle
the current value between the Table and Wall locations. Selection
of Buckys is indicated when the dark line is behind or below the
receptor surface. If the Bucky is not selected, the word "NONE" will
be shown. International symbols have been used to identify the
Table and Wall receptors.