Safety
VX65 User Guide
10 / 86
Declaration electromagnetic immunity
IMMUNITY test
IEC 60601 TEST
LEVEL
Compliance
level
Electromagnetic environment
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3V, 6V
10V/m
10V/m from 0.15
to 80MHz;
10V/m from 0.15
to 80MHz and
80% AM
at 1kHz
10V/m from
80MHz
to 2.7GHz
3V, 6V
10V/m
10V/m from
0.15 to
80MHz;
10V/m from
0.15 to
80MHz and
80% AM
at 1kHz
10V/m from
80MHz
to 2.7GHz
Portable and mobile RF
communications equipment should be
used no closer to any part of the VX65
including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
D Interference may occur in the vicinity
of equipment marked with the following
symbol:
Recommended separation distances between portable and mobile RF
communications equipment and the [ME EQUIPMENT or ME SYSTEM]
Rated
maximum
output
power of
transmitter
W
Separation distance according to frequency of transmitter m
150 kHz to 80
MHz
outside ISM
bands
150 kHz to 80
MHz
in ISM bands
80 MHz to 800
MHz
800 MHz to 2,5
GHz
0.01
0.12
0.2
0.4
1
0.1
0.37
0.64
1.3
2.6
1
1.17
2
4
8
10
3.7
6.4
13
26
100
11.7
20
40
80

Table of Contents