Instructions for care
Important!
➢ Before cleaning, make sure to disconnect the VX65 from the power supply!
➢ Do not spray or pour any liquid directly on the device.
➢ Do not use caustic or abrasive cleaners.
➢ Ensure that no moisture penetrates the system during cleaning and disinfection.
➢ Replace the paper pads after each patient.
Note: The selected disinfectant and method of disinfecting must comply with national
disinfecting regulations. The use of ethanol (DAB 10) 80 % vol. or isopropanol 70 % vol. is
recommended for disinfecting parts which are in direct contact with the patient.
Cleaning the Unit
Cleaning the surfaces
To clean the plastic housing, dampen a cloth with a commercial, non-abrasive cleaner and
gently wipe the top, bottom and front surfaces.
Cleaning the forehead rest
Regular cleaning of the forehead rest with a soft cloth and alcohol is strongly recommended.
The forehead rest is the only part of the machine that comes into contact with the patient.