Specifications
Chapter 1: Introduction
Electrical Description
Table 1-6. Electrical Specifications
Description Specification
Vo ltage 109–125V AC at 50–60 Hz
Wire—standard and emergency power Type AWT/MTW
Stranded copper
600V
Flame retardant
Heat resistant
Wire—power circuits/lighting circuits 12 AWG (color-coded
according to wiring dia-
gram)
Wire—ground conductor 12 AWG
Stranded copper
Green
Wire—raceway ground conductor 10 AWG to terminate at the
service chase ground bus.
Grounding and bonding Each receptacle must have
a grounding conductor
terminated at the raceway
grounding post next to the
service chase.
Page 1 - 10 Integris® 2001 Headwall System Service Manual (man148ra)