EasyManua.ls Logo

WittRock Integris 2001 - Low Voltage Controller Service

Default Icon
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4.10 Low Voltage Controller
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.10 Low Voltage Controller
Tools required: Adjustable wrench Screwdriver
Phillips head screwdriver
Removal
SHOCK HAZARD:
Locate the involved building circuit breaker panel. Set the involved
circuit breaker to OFF. Lock out and tag out the breaker. Failure to
follow this procedure could result in personal injury or equipment
damage.
1. Set the involved circuit breaker to OFF.
2. Lock out and tag the circuit breaker.
3. Gain access to the low voltage controller.
4. Remove the three wire nuts (E) from the wire joints leading to the raceway
(see figure 4-10 on page 4-23).
Page 4 - 22 Integris® 2001 Headwall System Service Manual (man148ra)

Table of Contents