2.5 Timed Light Switch and Toggle Switch Do Not Operate Properly
Chapter 2: Troubleshooting Procedures
2.5 Timed Light Switch and Toggle Switch Do Not Operate
Properly
SHOCK HAZARD:
Use care when checking live voltages. Do not touch live terminals,
wires, and ground. Failure to use caution could result in personal injury.
1. The building circuit breaker is in the ON position.
Yes No
Reset the breaker to the ON position.
2. A voltage of 109 to 125V AC is available between the lug and the terminal
of the breaker.
Yes No
Replace the circuit breaker according to the manufacturers
instruction.
3. A voltage of 109 to 125V AC is available at th
e wire terminals on the
toggle switch.
Yes No
Repair or replace the wiring between the circuit breaker and the
toggle switch.
4. Place the toggle switch in the ON position. The lights turn ON.
Yes No
Replace the applicable light bulb.
5. The light is operational.
Yes No
Replace the toggle switch. See Timed Light Switch and Toggle
Switch on page 4-17.
6. Place th
e toggle light switch in the OFF position. The lights turn OFF at the
setting of the timer (015 minutes).
Yes No
Adjust the timer to the correct setting. See Timed Light Switch
and Toggle Switch on page 4-17.
2
Integris® 2001 Headwall System Service Manual (man148ra) Page 2 - 13