2.5 Timed Light Switch and Toggle Switch Do Not Operate Properly
Chapter 2: Troubleshooting Procedures
7. The light turns off at the correct time.
Yes No
→ Replace the timer assembly. See Timed Light Switch and
To
ggle Switch on page 4-17. If this solves the problem, go
to Final Actions on page 2-5. Otherwise, call WittRock
Technical Support at (812) 222- 0373.
8. Go to Final Actions on page 2-5.
Page 2 - 14 Integris® 2001 Headwall System Service Manual (man148ra)