4.4 Line Voltage Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.4 Line Voltage Switch
Tools required: Screwdriver Phillips head screwdriver
Removal
SHOCK HAZARD:
Locate the involved building circuit breaker panel. Set the involved
circuit breaker to OFF. Lock out and tag out the breaker. Failure to
follow this procedure could result in personal injury or equipment
damage.
1. Set the involved circuit breaker to OFF.
2. Lock out and tag the circuit breaker.
3. Using the phillips head screwdriver, remove the faceplate screws (C) and
the faceplate (B) (see figure 4-4 on page 4-9).
Figure 4-4. Line Voltage Switch Removal
m148a014
4
Integris® 2001 Headwall System Service Manual (man148ra) Page 4 - 9