Getting Started
Chapter 2: Troubleshooting Procedures
Getting Started
WARNING:
Only facility-authorized personnel should
troubleshoot the Integris®
2001 Headwall System. Troubleshooting by unauthorized personnel
could result in personal injury or equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined
(each step assumes the previous step has been completed). In each step, the
normal operation of the product can be confirmed by answering Yes or No to
the statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one
component is listed, replace them in the given order.
To begin gathering information about the problem, start with Initial Actions.
To isolate or identify a problem and to verify the repair after completing each
corrective action (replacing or adjusting a part, seating a connector, etc.),
perform the Function Checks.
To verify the repair, perform the Final Actions after the Function Checks.
If troubleshooting procedures do not isolate the problem, call WittRock
Technical Support at (812) 222- 0373 for assistance.
2
Initial Actions
To gather information from operators concerning problems with the Integris®
2001 Headwall System, use Initial Actions. Note symptoms or other
information concerning the problem that the operator describes. This
information helps identify the probable cause.
1. Someone who can explain the problem is available.
Yes No
↓ → Go to “Function Checks” on page 2-4.
2. Ask that person to demonstrate or explain the problem. The problem can be
duplicated.
Yes No
↓ → Go to “Function Checks” on page 2-4.
3. The problem is a result of improper operator action.
Integris® 2001 Headwall System Service Manual (man148ra) Page 2 - 3