2.2 Dimmer Switch Does Not Operate the Light
Chapter 2: Troubleshooting Procedures
The resistance value drops to approximately zero.
Yes No
Replace the dimmer switch. See Dimmer Switch on page 4-7.
6. The dimmer switch operates the light.
Yes No
Repair or replace the wiring between the dimmer switch and the
light. If this solves th
e problem, go to Final Actions on page
2-5. Otherwise, call WittRock Technical Support at (812) 222-
0373.
7. Go to Final Actions on page 2-5.