2.3 Line Voltage Switch Does Not Operate Attached Device
Chapter 2: Troubleshooting Procedures
7. The line voltage switch is operative.
Yes No
↓ → Repair or replace the wiring between the switch and the wall box
o
r the switch and the Integris® 2001 Headwall System circuit
breaker, if equipped. If this solves the problem, go to “Final
Actions” on page 2-5. Otherwise, call WittRock Technical
Support at (812) 222- 0373.
8. Go to “Final Actions” on page 2-5.
2
Integris® 2001 Headwall System Service Manual (man148ra) Page 2 - 11