2.3 Line Voltage Switch Does Not Operate Attached Device
Chapter 2: Troubleshooting Procedures
7. The line voltage switch is operative.
Yes No
→ Repair or replace the wiring between the switch and the wall box
o
r the switch and the Integris® 2001 Headwall System circuit
breaker, if equipped. If this solves the problem, go to Final
Actions on page 2-5. Otherwise, call WittRock Technical
Support at (812) 222- 0373.
8. Go to Final Actions on page 2-5.