Communication Function
8-13
IM CA150E
8
SO Normal SettingStarts/stops and queries source
○ ○
Command = SOm<Delimiter>
→
Answer = SOm <Delimiter>
Command = SO?<Delimiter>
→
Return = SOm <Delimiter>
Parameter
m = 0: Stop/1: Start
SR Normal SettingSets and queries the source range
○ ○Command = SRm<Delimiter> → Answer = SRm <Delimiter>
Command = SR?<Delimiter>
→ Return = SRm <Delimiter>
Parameter
[DCV] m = 0: 100 mV/1: 1 V/2: 10 V/3: 30V
[DCA]
(When normal) m = 0: 20 mA/1: 4-20 mA
(When Calibrating) m = 0: 20 mA/2: 20 mA SINK
[Ω]
(When normal) m = 0: 500 Ω/1: 5 kΩ/2: 50 kΩ
(When Calibrating) m = 0: 500 Ω (1 mA)/1: 5 kΩ (0.1 mA)/
2: 50 kΩ (0.05 mA)/
3: 500 Ω (5 mA)/4: 5 kΩ (0.5 mA)
[TC] m = 0: K/1: E/2: J/3: T/4: R/5: B/6: S/ 7: N/8: L/9: U
[RTD] m = 0: PT100/1: JPT100
[PULSE] m = 0: 100 Hz/1: 1000 Hz/2: 10 kHz/3: 50 kHz/4: 1000 CPM
SY Normal SettingSwitches and queries normal mode and setting mode
○ ○Command = SYm<Delimiter> → Answer = SYm <Delimiter>
Command = SY?<Delimiter>
→ Return = SYm <Delimiter>
Parameter
m = 0: Normal Mode/1: Setting Mode
default Value
m = 0 (Normal Mode)
TE Normal SettingSets and queries TC and RTD (source) display
○ ×
Command = TEm<Delimiter>
→
Answer = TEm <Delimiter>
Command = TE?<Delimiter>
→
Return = TEm <Delimiter>
Parameter
m = 0: Temperature Value/1: mV Value (Resistance Value)/2: Room
Temperature
default Value
m = 0 (Temperature Value)
TT Normal SettingSets and queries the international temperature standard
× ○Command = TTm<Delimiter> → Answer = TTm <Delimiter>
Command = TT?<Delimiter>
→ Return = TTm <Delimiter>
Parameter m = 0: IPTS-68/1: ITS-90
default Value m = 1 (ITS-90)
8.3 Detailed Description of Commands