Specications
11-7
IM CA150E
Serial interface: RS232
Display: Segmented LCD
Backlight: LED backlight, auto off after 10 minutes
Warm-up time: Approximately 5 minutes
Power supply:
Six AA-size alkaline batteries
Dedicated NiMH battery or dedicated AC adapter (optional)
The charging time is approximately 6 hours.
Battery life:
When measurement is on and output is 5 V DC/10 kΩ or greater:
Approximately 8 hours for alkaline batteries and
10 hours for the dedicated NiMH battery
Auto power off: Approximately 10 minutes (can be disabled)
Insulation resistance: 50 MΩ or more at 500 V DC between the input
terminals and output terminals
Withstanding voltage:
350 V AC for 1 minute between the measurement
terminals and source terminals
Operating temperature/humidity ranges:
0 to 40°C, 20 to 80% RH (no condensation)
Storage temperature/humidity ranges:
-20 to 60°C, 90% RH or less (no condensation)
Dimensions: Approximately 251 × 124 × 70 mm
Weight: Approximately 1000 g (including batteries)
Accessories:
Source lead cables, measurement lead cables,
carrying case, fuse (spare), terminal adapter, alkaline batteries (six),
user's manual, ferrite core (two).
11. Specications