EasyManuals Logo

YOKOGAWA MuR10000 User Manual

YOKOGAWA MuR10000
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
6-15
IM 04P01B01-01E
Setup Operations for Convenient Functions (Setting Mode)
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Explanation
Setting Correction Points (Measured Values) and Correction Values
Using revise values
Calibration point
B4
A3
A2
A1
A4
B3
B2
B1
Scale value
Correction
value
Measured
value
Correction using revise values
Correction point (measured value) = A
Correction value = B – A
Using absolute values
Calibration point
B4
A3
A2
A1
A4
B3
B2
B1
Scale value
Correction
value
Measured
value
Correction using absolute values
Correction point (measured value) = A
Correction value = B
Example of a two-point correction
Calibration
Point
Corrected
Value
Measured
value (A)
Correction Value
Revise Value
(B – A)
Absolute
Value (B)
1 10.0°C 9.8°C 0.2°C 10.0°C
2 90.0°C 90.5°C – 0.5°C 90.0°C
Setup Conditions of Measured and Correction Values
Setting range of measured and correction values when using revise values
A1<A2≤A3≤...≤A16
(A1 + B1) < (A2 + B2) < (A3 + B3) < ... < (A16 + B16)
A and B are within the measurable range or scaling range of the range type*
* Within the scaling range: –5% to 105% of the scale value or –20000 to 30000 (the decimal
place is the same as the setting for the scale value)
Setting range of measured and correction values when using absolute values
A1<A2≤A3≤...≤A16
B1 < B2 < B3 < ... < B16
A and B are within the measurable range or scaling range of the range type*
* Within the scaling range: –5% to 105% of the scale value or –20000 to 30000 (the decimal
place is the same as the setting for the scale value)
Note
Make sure that the leftmost value is less than the rightmost value for span and scale.
Calibration correction cannot be set on ON/OFF input (DI), delta computation, and square
root computation channels. In addition, calibration correction cannot be used on all channels
if the bias function is enabled.
<Related Topics> Enabling the calibration correction function: Section 7.15
Setting the calibration correction function: Section 7.23
6.13 Performing Calibration Correction (/CC1 Option)

Table of Contents

Other manuals for YOKOGAWA MuR10000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MuR10000 and is the answer not in the manual?

YOKOGAWA MuR10000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMuR10000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals