Troubleshooting
10.1 A List of Error Messages
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Code Message Explanation/Countermeasures
351
This command cannot be specified in the current mode.
352 The option is not installed.
353
This command cannot be specified in the current setting.
354 This command is not available during calculation.
390 Command error.
391 Delimiter error.
392 Parameter error.
393 No permission.
394 No such connection.
395 Use “quit” to close this connection. Attempted to disconnect its own connection.
396 Failed to disconnect.
397 No TCP control block. The control block of the specified connection cannot be found.
400 Input username.
401 Input password.
402 Select username from ‘admin’ or ‘user’. If the recorder is configured not to use the user name and
password, use user names ‘admin’ or ‘user’.
403 Login incorrect, try again!
404 No more login at the specified level is acceptable.
420 Connection has been lost.
421
The number of simultaneous connection has been exceeded.
422 Communication has timed-out.
Warning Messages
Code Message Explanation/Countermeasures
600 Initialized. Settings and measured data have been initialized. If this
message always appears when you turn the power off and then
back on, contact your nearest YOKOGAWA dealer.
612 Some items not set. Some of the settings could not be loaded because the model,
the number of channels, options, the setup items in basic setting
mode, or the like was different.
If you change a setup item in Basic Setting mode from the
Configuration Software, load the settings of Basic Setting mode.
See “Loading the Settings of Basic Setting Mode and Setting
Mode”in“LoadingSetupData”oftheμR10000/μR20000SD
Memory Card (/EM1 Option) User’s Manual (IM 04P01B01-03E).
613 Less than 10% free space on the SD card.
The amount of free space on the SD memory card is low. Stop
recording to the SD memory card and replace the SD memory card.
614 Less than 25 MB free space on the SD card. The amount of free space on the SD memory card is low.
Replace the SD memory card.
System Errors
Code Message Explanation/Countermeasures
902 RAM failure. Contact your nearest YOKOGAWA dealer.
910 A/D error. Contact your nearest YOKOGAWA dealer.
921 A/D calibration value error. Contact your nearest YOKOGAWA dealer.
922 A/D calibration is in the wrong order. Contact your nearest YOKOGAWA dealer.
930 Memory acquisition failure. Contact your nearest YOKOGAWA dealer.
940 The ethernet module is down. Contact your nearest YOKOGAWA dealer.
950 A/D number error. Contact your nearest YOKOGAWA dealer.
951 EEPROM write error. Contact your nearest YOKOGAWA dealer.
960 Ribbon error Contact your nearest YOKOGAWA dealer.
961 Printer error Contact your nearest YOKOGAWA dealer.
962 Plotter error Contact your nearest YOKOGAWA dealer.
963 Pen 1 error Contact your nearest YOKOGAWA dealer.
964 Pen 2 error Contact your nearest YOKOGAWA dealer.
965 Pen 3 error Contact your nearest YOKOGAWA dealer.
966 Pen 4 error Contact your nearest YOKOGAWA dealer.
B5 (No messages) Contact your nearest YOKOGAWA dealer.
B6 (No messages) Contact your nearest YOKOGAWA dealer.

Table of Contents

Other manuals for YOKOGAWA MuR10000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MuR10000 and is the answer not in the manual?

YOKOGAWA MuR10000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMuR10000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals