11-7
IM 04P01B01-01E
Maintenance
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Applying the Changes and Returning to Operation Mode
Press the key or SHIFT + key to select End and then press the key. Press
the
key to select Store and then press the key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key to
return to the Basic= screen.
Note
When the Customized Menu is enabled, the operation is different.
When “P_Adj” in the Customized Menu is On
1. Perform steps 1-4 of “Adjusting the Pen Position.”
2. When the instrument enters Basic Setting mode, “Menu kind” appears. Select “Free”
then press a key.
3. Perform the steps starting with step 5 of “Adjusting the Pen Position.”
• When“P_Adj”intheCustomizedMenuisO󰀨
See “Disabling the Customized Menu” or “Changing the Settings without Disabling the Basic
Setting Mode Lock.”
<Related Topics> Enabling/Disabling the customized menu: Section 7.22
Changing the settings without releasing the basic setting mode lock:
Section 7.22
11.5 Adjusting the Pen Position (Pen Model)

Table of Contents

Other manuals for YOKOGAWA MuR10000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MuR10000 and is the answer not in the manual?

YOKOGAWA MuR10000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMuR10000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals