EasyManuals Logo

YOKOGAWA MuR10000 User Manual

YOKOGAWA MuR10000
272 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #255 background imageLoading...
Page #255 background image
12-19
IM 04P01B01-01E
Specifications
1
2
3
4
5
6
7
8
9
10
11
12
App
Index
Isolation
Item Specifications
Insulation resistance Eachterminaltogroundterminal:20MΩormore(at500VDC)
Dielectric strength Power supply to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Contact output terminal to ground terminal: 1500 VAC (50/60 Hz), 1 minute
Measuring input terminal to ground terminal: 1000 VAC (50/60 Hz), 1 minute
Between measuring input terminals: 1000 VAC (50/60 Hz), 1 minute (except for RTD input terminal)
Remote input terminal to ground terminal: 500 VDC, 1 minute
Grounding Be sure to set a low grounding resistance.
Transport and Storage Conditions
Item Specifications
Ambient temperature –25 to 60°C
Ambient humidity 5 to 95%RH (no condensation)
Vibration 10 to 60 Hz, 4.9 m/s
2
maximum
Shock 392 m/s
2
maximum (in packaged condition)
Supported Standards
Item Specifications
CSA Certified by CSA22.2 No. 61010-1, CSA C22.2 No.61010-2-030 (certified by NRTL/C
*
), overvoltage
category II
or I
1
, measurement category II
2
, pollution degree 2
3
* “US” (USA) and “C” (Canada) are indicated to the right and left of the CSA mark, respectively, to
show that NRTL is included.
CE
EMC EN61326-1 compliance, Class A, Table 2 (For use in industrial locations)
EN61000-3-2 compliance
EN61000-3-3 compliance
EN55011 compliance, Class A, Group 1
Low voltage
directive
EN61010-1, EN 61010-2-030 compliance, overvoltage category II
or I, measurement category II,
pollution degree 2
EMC regulatory arrangement in Australia and New Zealand
EN55011 compliance, Class A, Group 1
KC marking Electromagnetic wave interference prevention standard, electromagnetic wave protection standard
compliance
WEEE directive Compliant
1 Overvoltage category describes a number which defines a transient overvoltage condition. It implies the regulation
for impulse withstand voltage. “II” applies to electrical equipment which is supplied from the fixed installation like a
distribution board.
II: Applied to standard power supply (100-240 V AC)
I: Applied to /P1 option (21.6-26.4 V AC/DC)
2 Applies to measuring circuits connected to low voltage installation, and electrical instruments supplied with power from
fixed equipment such as electric switchboards.
3 “Pollution degree” describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface
resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution occurs.
12.6 General Specifications

Table of Contents

Other manuals for YOKOGAWA MuR10000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MuR10000 and is the answer not in the manual?

YOKOGAWA MuR10000 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMuR10000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals