12
2.2.2 Tubing Installation
All tubing assemblies used with the Multi-Channel Online Monitor are not compatible for vacuum
autoclave cycles. Do NOT use any vacuum autoclave cycles when autoclaving these items,
Vacuum cycles include pre-vacuum and post-autoclave vacuum drying cycles.
The Tubing Manifold cannot be autoclaved.
2.2.2.1 Peristaltic Pump Tubing
2.2.2.1.1 Pump Sample Tubing
1. Carefully remove the new peristaltic pump sample tubing (smaller diameter tubing) from its package.
2. Open the peristaltic pump by turning the release knob counterclockwise (Figure 2-4).
3. Connect the male Luer end of the pump sample tubing to the bottom fitting of the manifold (Figure 2-5).
Figure 2-4
4. Insert the pump sample tubing into the rear slot of the peristaltic pump.
5. Connect the female Luer end of the pump sample tubing to the 2900 Series Analyzer’s sample line connector.
See Section 2.2.4.3 Sample and Waste Line Installation, for sample line connections.