13
2.2.2.1.2 Pump Waste Tubing
1. Carefully remove the new peristaltic pump waste tubing (larger diameter tubing) from its package.
2. Insert the pump waste tubing into the front slot of the peristaltic pump (Figure 2-5). NOTE: Male Luer end on right
side of pump.
3. Verify the tubing is aligned with the V notches in the pump.
4. Close the peristaltic pump by turning the release knob clockwise firmly until it snaps closed. NOTE:
Ensure that the pump tubing stays in place as the pump is closed.
Figure 2-5
5. Connect the 2900 Series Analyzer’s waste line from the sample cup to the right end of the waste pump tubing.
See Section 2.2.4.3 Sample and Waste Line Installation, for waste line connections.
2.2.3 Reagent and Waste Bottle Installation
1. Remove the two reagent bottles and the waste bottle from their packaging.
2. Fill each reagent bottle with the recommended fluid(s) listed below. If a particular solution will not be used, i.e.,
antiseptic, there is no need to fill the bottle with the respective solution.
• Antiseptic Bottle (chemical disinfectant/sanitizer): 70% ethanol or 70% IPA are the authorized disinfectant
solutions that may be used.