If the Multi-Channel Online Monitor sample tubing is pinched or crimped, tubing damage may
occur during the autoclave cycle.
Figure 2-13
7. Autoclave the bioreactor with the sample probe and Multi-Channel Online Monitor sample tubing assembly.
Do NOT use any vacuum autoclave cycles when autoclaving the tubing
assembly. Vacuum cycles include pre-vacuum and post-autoclave vacuum
drying cycles.
8. After the autoclave cycle is complete, remove the vessel from the autoclave and allow it to cool to room temperature.
Tubing should be straight and not
pinched or crimped