80
F013: All Modules Deactivated From Not Being Able to Calibrate.
Ensure all probes are calibrated before starting Multi-Channel Online Monitor session.
F014: Unable to put YSI in Remote Mode.
Make sure that the 2900 Series Analyzer is in the Run screen before starting Multi-Channel Online Monitor
session.
F017: Module Busy. (Specific module is referenced in message).
Ensure probes are calibrated and that sampling is finished before starting Multi-Channel Online Monitor session.
F018: YSI System Status Busy.
Ensure probes are calibrated and that sampling is finished before starting Multi-Channel Online Monitor session.
F019: 2/3/4 Status Timeout for Module 1.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F020: Module 1 Disabled and No Other Modules Active.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F021: 2/3/4 Status Timeout for Module 2.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F022: Module 2 Disabled and No Other Modules Active.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F023: 2/3/4 Status Timeout for Module 3.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F024: Module 3 Disabled and No Other Modules Active.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F025: 5/6/7 Status Timeout for Module 1.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F026: 5/6/7 Status Timeout for Module 2.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F027: 5/6/7 Status Timeout for Module 3.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.
F028: Results Request Timeout.
Check status of Module on 2900 Series Analyzer. Run a manual sample with probe(s) and see if analysis is
completed.