Error Messages
F001: System is in CFR Mode.
Take YSI 2900 Series Analyzer out of 21 CFR part 11 mode and enter Run before starting remote sample
F002: Waste Bottle Full. (Specific bottle is referenced in message).
Empty waste bottles connected to active modules. Rinse and dry waste bottle cap.
If Bottle Check is enabled on Multi-Channel Online Monitor, make sure that all bottles are enabled in Settings >
Fluid Detection on the 2900 Series Analyzer.
F003: Invalid Waste Bottle ID. (Specific bottle is referenced in message).
Check RS232 communications cable between Multi-Channel Online Monitor and 2900 Series Analyzer.
Potential failure of Multi-Channel Online Monitor or the YSI 2900 Series Analyzer. Review audit trail to diagnose
problem.
F004: Buffer Bottle Empty for Module. (Specific bottle is referenced in message).
Fill configured buffer bottle with appropriate buffer.
If Bottle Check is enabled on Multi-Channel Online Monitor, make sure that all bottles are enabled in Settings >
Fluid Detection on the 2900 Series Analyzer.
F005: Code 4. Invalid Buffer Bottle ID. (Specific bottle is referenced in message).
Check RS232 communications cable between Multi-Channel Online Monitor and 2900 Series Analyzer.
Potential failure of Multi-Channel Online Monitor or the YSI 2900 Series Analyzer. Review audit trail to diagnose
problem.
F006: Calibrator Bottle Empty for Probe. (Specific probe is referenced in message).
Fill configured calibration bottle with appropriate buffer.
If Bottle Check is enabled on Multi-Channel Online Monitor, make sure that all bottles are enabled in Settings >
Fluid Detection on the 2900 Series Analyzer.
F007: Invalid Calibrator Bottle ID. (Specific bottle is referenced in message).
Check RS232 communications cable between Multi-Channel Online Monitor and 2900 Series Analyzer.
Potential failure of Multi-Channel Online Monitor or the 2900 Series Analyzer. Review audit trail to diagnose
problem.
F008: All Probes/Modules are Disabled.
Ensure that all required calibration bottles and buffer bottles are not empty. Verify that fluid detection is Enabled for
all bottles in the 2900 Series Analyzer.
F009: Probe selected on Multi-Channel Online Monitor is disabled on 2900 Series Analyzer. (Specific probe is referenced
in message).
Either unselect the probe in the Multi-Channel Online Monitor Chemistry Configuration or enable the probe on the
2900 Series Analyzer.
F010: Probe selected on Multi-Channel Online Monitor is invalid for 2900 Series Analyzer. (Specific probe is referenced in
message).
Unselect the probe in the Multi-Channel Online Monitor Chemistry Configuration.
F011: Unable to Put YSI in Remote Mode
Place the 2900 Series Analyzer in the Run screen starting Multi-Channel Online Monitor session.
F012: Could Not Calibrate Module. (Specific module is referenced in message).
Ensure all probes are calibrated before starting Multi-Channel Online Monitor session.