Restarting the Device
Certain events require the E Series products to be
restarted after they shut off or become inoperative.
One example is when the battery runs down and the unit
shuts off. In this case, perform these steps in this order:
1. Turn the selector switch to the OFF position
2. Remove the battery.
3. Insert a new battery.
4. Turn the selector switch to the desired operating
mode to resume operation.
This sequence is necessary to restart the device, and
can also be used to clear some X FAULT XX messages,
when immediate use of the device is required.
Note that some settings (for example, alarm settings,
lead selection, ECG size) may need to be restored from
their default values when operation is resumed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL E Series and is the answer not in the manual?

ZOLL E Series Specifications

General IconGeneral
Display TypeColor LCD
ECG MonitoringYes
PacingYes
Pacing Rate30-180 ppm
SpO2Optional
NIBPOptional
EtCO2Optional
PrinterIntegrated
Data ManagementYes
CPR FeedbackYes
IP RatingIP55
Pacing Output0 - 140 mA
Battery TypeRechargeable lithium ion
Battery Life4 hours
Energy Selection1-360 Joules
Display Size6.5 inches