Advisory Defibrillation
9650-1210-01 Rev. R 4-5
and does not require depression of the ANALYZE
button for operation.
• CHECK PADS / POOR PAD CONTACT
The MFE Pads are no longer properly attached to the
patient or the cable connections have become loose.
Check that the MFE Pads are making good contact
with the patient’s skin and that the cables are all
securely connected. This voice prompt is issued only
if MFE pads were previously connected to the patient.
Troubleshooting
If your E Series unit does not function as expected, see
the Defibrillator Troubleshooting section starting on
page 13-5.