9650-1210-01 Rev. R 9-1
SECTION 9
NONINVASIVE TEMPORARY PACING
(PACER VERSION ONLY)
Noninvasive Temporary Pacing
Some E Series products contain a VVI demand
pacemaker — a safe and effective design for
Noninvasive Temporary Pacemakers. Proper demand
pacing requires a reliable, high-quality surface ECG
signal.
Determine patient condition and provide care
following medical protocols.
Prepare Patient
Remove all clothing covering the patient’s chest. Dry
chest if necessary. If the patient has excessive chest
hair, clip it to ensure proper adhesion of electrodes.
1 Apply Electrodes and/or MFE Pads
Connect the electrodes to the ECG cable leads, and the
ECG cable to the E Series unit. Apply ECG electrodes
(refer to “ECG Monitoring” on page 10-1). Adjust the
ECG size and lead for a convenient waveform display.
Verify proper R-wave detection. The heart-shaped
symbol flashes with each R-wave when proper detection
is taking place.
Attach and connect the MFE pads
This is described in “MFE Pad Application/Connection”
on page 1-9.
2 Turn Selector Switch to PACER
Set Pacer Output to 0 mA
If the unit has just been turned on, the PACER OUTPUT
is automatically set to 0 mA.
3 Set Pacer Rate
Set PACER RATE to a value 10-20 ppm higher than
patient’s intrinsic rate. If no intrinsic rate exists, use
100 ppm.
Paddles are a defibrillation-protected Type BF patient connection.
ECG leads are a defibrillation-protected Type CF patient connection.
WARNING
To avoid risk of electrical shock, do not touch the gelled area of the MFE Pads while pacing.
MFE Pads should be replaced after 8 hours of continuous pacing (2 hours for Radiolucent stat-padz) to ensure
maximum patient benefit.
Pacing with CPR-D-padz is not recommended. For the most effective pacing, use MFE pads.
Prolonged pacing (in excess of 30 minutes), particularly in neonates or adults with severely restricted blood flow,
may cause burns. Periodic inspection of the underlying skin is recommended.
If the unit was not
turned off and less than 10 minutes have elapsed since the pacing mode was last used,
reactivating the pacer mode causes pacing to immediately resume at the previously selected mA and ppm
settings.
Pacing must be turned off before defibrillating with a second defibrillator. Otherwise, the E Series unit may be
damaged.
0!#%2
/&&
-/.)4/2
$%&)"

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL E Series and is the answer not in the manual?

ZOLL E Series Specifications

General IconGeneral
Display TypeColor LCD
ECG MonitoringYes
PacingYes
Pacing Rate30-180 ppm
SpO2Optional
NIBPOptional
EtCO2Optional
PrinterIntegrated
Data ManagementYes
CPR FeedbackYes
IP RatingIP55
Pacing Output0 - 140 mA
Battery TypeRechargeable lithium ion
Battery Life4 hours
Energy Selection1-360 Joules
Display Size6.5 inches