E Series Service Manual
1-20 9650-1450-01 Rev. D
11.0 Defibrillator Self Test
WARNING!
SHOCK HAZARD!
TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY BEFORE YOU START
CONDUCTING THE DEFIBRILLATOR TESTS.
Keep hands and all other objects clear of the multi-function cabl
e connections and defibrillator analyzer when discharging
the defibrillator.
Before you discharge the defibrillator, warn everyone
near the equipment to STAND CLEAR.
Caution Do NOT internally discharge the unit more than 3 times in 1 minute. Note that the rapid repetition of internal discharges at more than 30
Joules may damage the unit.
Tools Needed:
• MFC Test Port Connector 1004-0149 with universal cable.
• MFC Test Adaptor Connector (Fluke Biomedical Part Number 3010-0378 or equivalent).
• QED 6 Defibrillator Analyzer.
• ECG Cable.
• Stop watch.
Test Setup:
1. Ensure the unit is turned off
2. Ensure the ECG cable is connected to the unit and analyzer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL E Series and is the answer not in the manual?

ZOLL E Series Specifications

General IconGeneral
Display TypeColor LCD
ECG MonitoringYes
PacingYes
Pacing Rate30-180 ppm
SpO2Optional
NIBPOptional
EtCO2Optional
PrinterIntegrated
Data ManagementYes
CPR FeedbackYes
IP RatingIP55
Pacing Output0 - 140 mA
Battery TypeRechargeable lithium ion
Battery Life4 hours
Energy Selection1-360 Joules
Display Size6.5 inches

Related product manuals