7.0 Heart Rate Display Test
Tools Needed:
ECG Cable (3 or 5 leads).
QED 6 Defibrillator Analyzer.
Test Setup:
1. Turn the selector switch to MONITOR. Press LE
AD button until “I” displays.
2. Connect the ECG leads to the QED 6 Defibrillator Analyzer.
3. Connect the ECG cable to the unit.
Do this... Verify that... Pass/Fail/NA
7.1
o o o
Set the QED 6 Defibrillator Analyzer to
120BPM.
The Heart Rate displays as 120 +/- 2 bpm
Record your results on the Mai
ntenance Tests Checklist.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL E Series and is the answer not in the manual?

ZOLL E Series Specifications

General IconGeneral
Display TypeColor LCD
ECG MonitoringYes
PacingYes
Pacing Rate30-180 ppm
SpO2Optional
NIBPOptional
EtCO2Optional
PrinterIntegrated
Data ManagementYes
CPR FeedbackYes
IP RatingIP55
Pacing Output0 - 140 mA
Battery TypeRechargeable lithium ion
Battery Life4 hours
Energy Selection1-360 Joules
Display Size6.5 inches