Page #183 background image
Operating Controls and Indicators
9650-1210-01 Rev. R 2-11
Note: You cannot add a patient name to summary
report events already stored in memory. The
patient name is stored only with those summary
events saved after the patient name has been
entered.
Modifying a Patient Name and ID
Number
To change an existing patient name and identification
number:
1. Press ID # softkey to display the Patient Name and ID
screen and keypad.
2. Use the scroll keys on top of the unit to select the
Next key (>>>) or the Prev key (<<<) on the keypad,
then press the Commit ( ) key, as many times as
necessary to move the cursor to the desired position.
3. Use the scroll keys on top of the unit to select the
Backspace key () on the keypad, then press the
Commit ( ) key to erase the selected character.
Repeat as necessary
4. Enter the new characters in the Patient Name field,
using the scroll keys on top of the unit to select
characters from the keypad and pressing the
Commit ( ) key to enter the selection.
5. When you have modified the patient name, move the
cursor to the Enter ( ) key on the keypad and press
the Commit ( ) key on top of the unit.
6. Repeat steps 2 through 5 to modify the Patient ID
field. Otherwise, press the Commit ( ) key again,
with the Enter ( ) key highlighted (or press the
Return softkey), to return to the main menu.
Erasing Summary Report Memory
To erase all stored information, press and hold the
SUMMARY button for approximately 4 seconds. Then
press the corresponding softkey to erase summary,
erase trend, or erase all event reports. An ERASING
REPORT message appears on the display.
Turning the unit off for more than 15 minutes, unless
configured otherwise, also erases summary and trend
report memory. All events remain stored until you erase
them or the unit has been turned off for a user-
configurable time period of 5 minutes to 36 hours.
ReturnID # Cancel
J
w
‹‹‹ ›››
PT. NAME: JOHN SMITH
PT. ID: 1234567
ReturnID #
Cancel
J
w
PT. NAME: J SMITH
PT. ID: 1234567
ReturnID # Cancel
J
w
PT. NAME: JANE SMITH
PT. ID: 1234567
#(!2'%2/.
35--!29
#/$%
-!2+%2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ZOLL E Series and is the answer not in the manual?

ZOLL E Series Specifications

General IconGeneral
Display TypeColor LCD
ECG MonitoringYes
PacingYes
Pacing Rate30-180 ppm
SpO2Optional
NIBPOptional
EtCO2Optional
PrinterIntegrated
Data ManagementYes
CPR FeedbackYes
IP RatingIP55
Pacing Output0 - 140 mA
Battery TypeRechargeable lithium ion
Battery Life4 hours
Energy Selection1-360 Joules
Display Size6.5 inches