E Series Operator’s Guide
A-14 9650-1210-01 Rev. R
Table A-1. Delivered Energy at Every Defibrillator Settings into a Range of Loads
Figures A-1 through A-20 show the rectilinear biphasic waveforms that are produced when the E Series defibrillator is
discharged into loads of 25, 50, 75, 100, 125, 150, and 175 ohms at each energy setting (200, 150, 120, 100, 85, 70,
50, 30, 20, 15, 10, 9, 8, 7, 6, 5, 4, 3, 2 and 1 joule[s]).
The vertical axis shows the current in amperes (A); the horizontal axis shows the duration in milliseconds (ms).
Selected
Energy
Load
25 50 75 100 125 150 175 Accuracy
1 J 1 J1 J1 J1 J1 J1 J1 J ±3J
2 J 1 J2 J3 J2 J2 J2 J2 J ±3J
3 J 2 J3 J4 J4 J3 J3 J3 J ±3J
4 J 3 J4 J5 J5 J5 J4 J 4 J ±3J
5 J 4 J5 J6 J6 J6 J5 J5 J ±3J
6 J 4 J6 J 8 J7 J7 J6 J6 J ±3J
7 J 5 J7 J9 J9 J8 J8 J7 J ±3J
8 J 6 J 8 J 10 J 10 J 9 J 9 J 8 J ±3J
9 J 7J 9 J 11 J 11 J 10 J 10 J 9 J ±3J
10 J 7 J 10 J 13 J 12 J 12 J 11 J 10 J ±3J
15 J 11 J 15 J 19 J 19 J 17 J 16 J 15 J ±3J
20 J 15 J 19 J 25 J 25 J 23 J 22 J 20 J ±3J
30 J 22 J 29 J 38 J 37 J 35 J 32 J 30 J ±15%
50 J 37 J 48 J 63 J 62 J 58 J 54 J 50 J ±15%
70 J 53 J 69 J 90 J 89 J 83 J 77 J 71 J ±15%
85 J 65 J 86 J 112 J 110 J 103 J 96 J 89 J ±15%
100 J 74 J 97 J 126 J 125 J 116 J 108 J 100 J ±15%
120 J 88 J 116 J 151 J 149 J 139 J 129 J 120 J ±15%
150 J 110 J 145 J 188 J 186 J 174 J 161 J 150 J ±15%
200 J 146 J 214 J 238 J 234 J 213 J 195 J 179 J ±15%