iQC Codes A-3
404F Temperature Error Rotor temperature too variable or outside
proper range during run. If symptom
persists, contact Abaxis Technical Service.
4042 Temperature Error Temperature deviated more than one degree
during run. Accuracy of certain chemistries
may be affected. Rerun with new rotor.
4043 Sample Blank Bead Error Sample blank bead appears turbid, possibly
due to sample quality. Confirm sample
integrity and rerun with new rotor.
4044 iQC Bead Mix Error Analyzer cannot determine exposure status
if iQC beads do not mix correctly. Rerun
with new rotor.
4045 RQC Error Reagent QC beads recovered below limit.
Rotor may have been stored incorrectly.
Rerun with new rotor.
4051 Flash Lamp Level Error Analyzer checked lamp (to make sure it
flashes when optical cuvette is sampled) but
detectors did not recognize lamp flashes. If
symptom persists, contact Abaxis Technical
Service.
4053 Engine Reset Error Analyzer component restarted unexpectedly.
Contact Abaxis Technical Service.
4069 Sample Quality Error Sample may contain physical interferents.
Confirm sample integrity and rerun rotor.
410D
410E
Bar Code Format Error Analyzer could not read rotor's bar code. If
symptom persists, contact Abaxis Technical
Service.
410F
4110
4111
Bar Code Format Error Analyzer did not recognize rotor. Contact
Abaxis Technical Service.
iQC Code Probable Cause Solution
Note: If the iQC code is not listed or if you cannot correct a particular problem, record the iQC num-
ber or print the report, and contact Abaxis Technical Service, 24 hours a day, 7 days a week,
at 1-800-822-2947, or send a fax to 1-877-900-9333.