EasyManuals Logo

ACUTRONIC fabian +nCPAP evolution User Manual

ACUTRONIC fabian +nCPAP evolution
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #125 background imageLoading...
Page #125 background image
17 Guide to Volume Guarantee (VG) 117
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
Step 3
Adjust thePinsp” setting until the measured VTe (on the right side of the display) reaches the
set Vguarant.
Figure 17-2: Set Pinsp parameter
IMPORTANT: To ensure the VG function works appropriately, a small plateau pressure in the
Pressure wave needs to be present. Plateau pressure means the patient lung pressure and the
circuit pressure reached the equilibrium state and are stabilized. Sufficient I-Time and I-Flow are
required for the inspiratory pressure to rise to the selected Pinsp and form a small plateau
pressure, DO NOT set the inspiratory time to a value too short to allow the plateau pressure to
be reached.
Step 4
Set in the Pmax area, the maximal pressure value to which the VG function can regulate. Pmax
should be set somewhere between Pinsp setting and High PIP alarm setting.
NOTE: To avoid unintentional breath abortions by reaching / exceeding High PIP, make sure that
the High PIP alarm is set to at least +2mbar above Pmax.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals