6 2 Warnings, cautions and notices
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
WARNING: Always unplug the ventilator from the power source before
opening the housing.
WARNING: NEVER use anti-static or electrically conductive tubing.
NOTE: To support patient and operator safety, the fabian HFO does not
contain VOCs, CO, CO
2
, O
3
, and particulates emitted above hazardous
thresholds. Acute, subacute/subchronic, and chronic toxicity from the
exposure to these compounds from the intended use of this device are not
expected.
WARNING: The device can only be isolated from the mains power supply by
removing the power cord completely.
• Ensure the power socket is always accessible for disconnection.
• DO NOT disconnect the power cable unless for Service purposes or
transport.
WARNING: DO NOT modify the equipment.
WARNING: Before applying non-original accessories, ensure that they are
biocompatible. All accessories supplied by ACUTRONIC Medical Systems for
use on fabian ventilators are biocompatible.
WARNING: When connected to a patient DO NOT simultaneously touch the
external power supply cord and the flow sensor connector cable.
WARNING: if the volume of the acoustic alarms is set less than the ambient
noise level this might impede an operator to recognize alarm conditions.
WARNING: NEVER cover the ventilator while in use.
WARNING: DO NOT position the ventilator in such a way that this adversely
affects its performance or makes it difficult to disconnect the ventilator from
the mains supply.
In case of emergency, removal of the mains plug from the wall outlet
disconnects the ventilator from mains power.
WARNING: in case of ventilator failure, the lack of immediate access to
appropriate alternative means of ventilation can result in patient death.
WARNING: Ensure that alarms are appropriately set before use of ventilator
on a patient.
WARNING: In case portions of the gas pathways through the VENTILATOR
become contaminated with body fluids or expired gases, immediately
contact ACUTRONIC Medical Systems.
WARNING:
• When selecting the neonatal patient size, a Neonatal Flow sensor should
be used.
• When selecting the pediatric patient size, a Pediatric Flow sensor should
be used.