64 7 Alarms
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
O
2
value out
of range
Sensor alarms
O
2
sensor or mixer
defective.
Repeat Calibration
and/or Replace O
2
sensor.
Disconnect the
device from AC
power and contact
your local
Acutronic Medical
Systems
Distributor.
I
4
SpO
2
module
disconnected
Sensor alarms
SpO
2
module is not
connected
Connect SpO
2
module
I
5
SpO
2
sensor
failure
Sensor alarms
SpO
2
sensor is faulty
Replace sensor or
Contact your local
ACUTRONIC Medical
Systems Distributor.
I
5
check SpO
2
sensor
Sensor alarms
SPO
2
sensor is off
patient or
malfunctioning
Connect SpO
2
sensor
to patient
I
5
Apnea alarm
Patient alarms
No spontaneous
patient breathing
detected.
Check patient,
Switch to
controlled
ventilation.
I
6
low SIQ
Patient alarms
SPO
2
sensor has low
SIQe (Signal Quality)
Check SpO
2
sensor for
correct contact with
patient or replace
sensor
I
7
FiO
2
at max
Patient alarms
FiO
2
has reached
maximum setting
Adjust settings
I
7
Lower
minimum FiO
2
Patient alarms
Lower minimum FIO
2
than set for PRICO
Adjust settings
I
7
Low Pmean
Patient alarms
Leak or
disconnection.
Check patient
connection.
Check tubing
system for leaks.
Check setting of
Pmean.
I
7
High Pmean
Patient alarms
Pressure rise in tubing
system.
Check patient,
Check tubing
system,
Check exhalation
valve.
Replace patient
system.
I
7