7 Alarms 65
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
High PIP
Patient alarms
Pressure rise in tubing
system; mechanical
inspiration was
reduced to relieve the
system.
NOTE: High PIP alarm
should be set to at
least +2mbar above
Pmax.
Check patient,
Check tubing
system.
Replace patient
system.
I
7
High PEEP
Patient alarms
PEEP 6 mbar above
set value for at least
15 seconds.
Check patient,
Check patient
connection,
Check tubing.
Adjust settings
I
7
High SpO
2
Patient alarms
SpO
2
measured
above set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
Low SpO
2
Patient alarms
SpO
2
measured below
set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
High Pulse
Rate
Patient alarms
Pulse rate measured
above set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
Low Pulse
Rate
Patient alarms
Pulse rate measured
below set limit
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8
Low Perfusion
Index
Patient alarms
Low SpO
2
perfusion
index
Check patient,
Check SpO
2
sensor
for correct contact
with patient
Adjust settings
II
8

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian +nCPAP evolution and is the answer not in the manual?

ACUTRONIC fabian +nCPAP evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian +nCPAP evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals